- TOR-80 is designed to measure the dielectric strength of transformer oil and other dielectric fluids according to IEC 60156, ASTM D877 and ASTM D1816
- Can work autonomously or transfer results to a PC, create a local database, generate reports, plot graphs and print results
- There is software for the implementation of standard and custom measurement procedures
- Max output voltage: up to 80 kV, sinusoidal
- Output voltage measurement accuracy ± 2 %
- Voltage increase rate: 0,1 to 5 kV/sec (adjustable)
- Resolution of output voltage indication: 100 V
- High voltage timeout after breakthrough: 4 ms typical (10 ms max)
- Operating voltage: 85 – 264 V a.c. (48…63 Hz)
- Power requirement: 250 VA max
- Measuring cell volume for oil: 500 cm3
- Temperature range of testing oil: 0…100 °C
- Unit operating conditions:
- Temperature: 0…50 °С
- Storage temperature: -20… 60 °С
- Relative humidity: Up to 90 % (without condensation)
More about breakdown voltage tester TOR-80 on GlobeCore website [EN]