- To measure air and neutral gases pressure in range -1000...1000 Pa
- Configuration with switches or software: measuring ranges, units, outputs settings
- Housing with simplified mounting system
- Optional with display or without display
- Range: -1000...1000 Pa
- Accuracy: ± 1.5 % of reading ± 3 Pa
- Resolution: 1 Pa; 0.1 mmH2O; 0.01 mbar; 0.01 inWG; 0.01 mmHG; 0.1 daPa; 0.001 kPa
- Measurement units: Pa, mmH2O, inWG, mmHG, daPa, kPa, hPa, mbar
- Response time: 0.3 s
- Output (Optional): 0-10 V ir 4-20 mA: 4-20 mA
- Overpressure tolerated: 21 000 Pa
- Conditions of use: 0…+50 °C. In non-condensing condition
- Communication to PC: USB-mini DIN cable
- Protection: IP65
- Power supply (Optional): 24 V AC / V DC: 16 / 30 V DC
More on pressure transmitter CP 112 on Sauermann website [EN]
CP 110-5 datasheet [EN]