This website uses cookies in order to provide better service. By using this website further you accept with this policy.
Product groups
Temperature (thermocouple) datalogger M1200

Temperature (thermocouple) datalogger M1200

Ask Product Code: CSM1200
Availability: Delivery in 1 - 2 weeks

COMET SYSTEM product ovrerview, PDF

COMET SYSTEM catalogs, PDFs

COMET SYSTEM videos on YouTube

  • Universal datalogger with 4 hardware MiniDIN inputs for thermocouples K, J, S, B, T, N.
  • Cold junction compensation (CJC).
  • Sensors of physical quantities with bipolar voltage output can also be connected. Two voltage ranges: -18...+18 mV and -60...+140 mV.
  • Multilogger can be installed permanently to Ethernet network or work as a portable device.
  • Measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.
  • Calibration certificate included.
  • 3 rechargeable AA batteries with a power adapter, wall holder and USB cable are included in the delivery.
  • Thermocouples are ordered separately.
  • Open collector transistor alarm output.

  

Operating temperature

-10...+60 °C

Measuring interval

1 s...24 h

Accuracy without probes - thermocouples K (NiCr-Ni)

±(0.3% from reading + 1.5°C) from -200 to +1300°C

Accuracy without probes - thermocouples J (Fe-Co)     

±(0.3% from reading + 1.5°C) from -200 to +750°C

Accuracy without probes - thermocouples S (Pt10%Rh-Pt)      

±(0.3% from reading + 1.5°C) from 0 to +1700°C

Accuracy without probes - thermocouples B (Pt30%Rh-Pt)

±(0.3% from reading + 1°C) from 100 to +1800°C,

uncompensated cold junction

Accuracy without probes - thermocouples T (Cu-CuNi)

±(0.3% from reading + 1.5°C) from -200 to +400°C

Accuracy without probes - thermocouples N (NiCrSi-NiSiMg)

±(0.3% from reading + 1.5°C) from -200 to +1300°C

Resolution

Better than 0.1°C

Accuracy - DC voltage from -60mV to +140mV

± 100uV, resolution better than 1uV

Accuracy - DC voltage from -18mV to +18mV

± 20uV, resolution better than 0.5uV

Memory capacity (approx.)

1 000 000 values (non-cyclic), 600 000 values (cyclic)

IP protection

IP20

Communication interface

Ethernet, USB, RS232

Ethernet communication protocols

WWW, SNMPv1, XML, DATALINK

Alarm protocols

E-Mail

Alarm output

OC, 100 mA, 24 V

Configuration

Comet Vision, WWW configuration

Power

AC adapter 230 VAC / 5 VDC (necessary for Ethernet interface)

NiMH rechargeable batteries 3 x 1.2 V AA (backup or portable application)

Alkaline batteries 3 x 1.5 V AA (for portable applications)

Dimensions

178 x 95 x 37 mm

 

More on recorder M1200 on Comet System website [EN]

M1200 datasheet [EN]